The conceptual design of the instrument is driven by the WEAVE science requirements, as defined by the Science Team, and translated to technical requirements by the Project and Instrument Scientists.  At the time of writing the requirements call for a dual-beam spectrograph  with the following characteristics:

Specification RequirementGoal
Field of view diameter 2 degrees 2 degrees  
MOS multiplex 800 1000 
Atmospheric dispersion compensation 0.4-1.0 µm, 50oZD 0.37-1.0 µm, 60oZD 
MOS fibre aperture >=1.2" 1.5" 
Open shutter efficiency (1  hour observations)70% 90% 
Wavelength coverage 400-950nm 370-1000nm 
Spectral resolution (full simultaneous coverage)5000 5000 
Spectral resolution (reduced coverage) 20000 20000 
Stray light as function of faintest target <1%  
Blue system efficiency 20% 25% 
Red system efficiency 25% 30% 
Multi-IFU multiplex 10 30 
Multi-IFU FOV 9" x 9" 9" x 12" 
Multi-IFU spaxel size 0.8" >1.2" 
Single IFU FOV3' x 3'  
Single IFU spaxel size 2.5" Inner 1", outer 2.5" 